Designed for biobanks and cell therapy facilities, the Arkylla stainless steel vapor-phase LN₂ storage system provides secure long-term preservation of valuable cell therapy products and biological samples. Its vacuum-jacketed stainless steel vessel with high-vacuum insulation delivers outstanding thermal performance. A microprocessor-controlled temperature and level monitoring system provides real-time tracking of high and low temperature points and liquid level, while a dual-solenoid automatic fill system ensures safe and reliable LN₂ supply. Vapor-phase storage eliminates direct contact with liquid nitrogen, minimizing cross-contamination risks, with the storage zone top temperature maintained stably below -190 °C—meeting long-term requirements for products including CAR-T, stem cells, and iPSCs. Wi-Fi-enabled remote monitoring and cloud-based data management support centralized oversight.
Vapor-phase storage eliminates direct contact with liquid nitrogen, minimizing cross-contamination risks
Vapor-phase top temperature maintained as low as -190 °C, meeting long-term storage temperature requirements
Dual-solenoid automatic fill system with primary and backup solenoids, supports manual emergency fill during power outages
Hot gas bypass design purges warm nitrogen from fill lines prior to dispensing, preventing temperature fluctuations that could compromise sample integrity
High-precision platinum temperature probe monitoring with ±1 °C display accuracy
10-inch touchscreen displays temperature curves, liquid level, and daily LN₂ consumption
One-Touch Defogging for quick sample location and retrieval
Wi-Fi-enabled remote monitoring and cloud-based data management
Critical operational data stored locally for 10 years and exportable via USB
Integrated folding step and auxiliary work platform for safe, convenient operation
Vacuum-jacketed stainless steel construction delivers excellent insulation, reducing LN₂ consumption
| Model | Arkylla-370 | Arkylla-460 | Arkylla-550 | Arkylla-800 | Arkylla-850 | Arkylla-1000 | Arkylla-1800 |
| Useful Volume (L) | 370 | 460 | 588 | 800 | 890 | 1031 | 1854 |
LN₂ Volume Under Platform (L) | 55 | 55 | 125 | 133 | 165 | 165 | 295 |
| Neck Opening Diameter (mm) | 320 | 320 | 445 | 445 | 465 | 465 | 635 |
| Internal Usable Height (mm) | 708 | 914 | 590 | 840 | 765 | 889 | 906 |
| Internal Diameter (mm) | 758 | 758 | 1012 | 1012 | 1113 | 1130 | 1467 |
| External Diameter (mm) | 820 | 820 | 1090 | 1090 | 1190 | 1190 | 1565 |
| Vessel Height (mm) | 1400 | 1600 | 1400 | 1400 | 1600 | 1700 | 1750 |
| Empty Weight (kg) | 212 | 264 | 310 | 310 | 400 | 495 | 805 |
| Full Weight (kg) | 513 | 638 | 750 | 750 | 1112 | 1319 | 2270 |
| 2 mL Cryovial Capacity | 16800 | 21000 | 28000 | 28000 | 42900 | 51000 | 95250 |
| No. of Racks | 12 | 12 | 24 | 24 | 32 | 32 | 60 |
| Auto Fill | Standard | Standard | Standard | Standard | Standard | Standard | Standard |
| One-Touch Defogging | Standard | Standard | Standard | Standard | Standard | Standard | Standard |
| Data Logging / USB / RS485 | Standard | Standard | Standard | Standard | Standard | Standard | Standard |
| Wi-Fi | Standard | Standard | Optional | Optional | Optional | Optional | Optional |